Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Amprobe ACD-10: (1 year) CAL VER /5520 DATE: 27-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 7 NUMBER OF LINES: 75 CONFIGURATION: Fluke 5520A STANDARD: Fluke 5500A/COIL ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Fluke 5500A/COIL 1.002 ASK- R P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: NONE 1.005 DISP Temperature: 18 - 28 degC 1.005 DISP Humidity: <80% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD {AC VOLTAGE} 1.008 DISP Set the FUNCTION switch to AC Volts. 1.009 DISP Connect the 5520 to the UUT as follows: 1.009 DISP [32] 5520A NORMAL HI to UUT V/OHM 1.009 DISP [32] 5520A NORMAL LO to UUT COM 1.010 HEAD AC VOLTAGE: {400V Range} 1.011 5520 400 300.0V 1.2% 0.3U 60H SI 2W 2.001 HEAD AC VOLTAGE: {700V Range} 2.002 5520 750 700V 1.2% 3U 60H SI 2W 3.001 5520 * S 3.002 HEAD {RESISTANCE} 3.003 DISP Connect the 5520 to the UUT as follows: 3.003 DISP [32] 5520A AUX HI to UUT V/OHM 3.003 DISP [32] 5520A AUX LO to UUT COM 3.003 DISP [32] 5520A NORMAL HI to UUT V/OHM 3.003 DISP [32] 5520A NORMAL LO to UUT COM 3.003 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 3.003 DISP the NORMAL leads to the SENSE leads AT the UUT. 3.004 DISP Set the FUNCTION switch to Ohms. 3.005 5520 0.0Z S CW 3.006 EVAL Is an audible tone present{(Continuity buzzer OK)}? 4.001 HEAD RESISTANCE: {40 kOhm Range} 4.002 5520 40 10.00kZ 2% 0.05U CW 5.001 5520 * S 5.002 HEAD {AC CURRENT} 5.003 DISP Set the FUNCTION switch to AC Amps. 5.004 DISP Connect a Fluke 5520A/COIL to the 5520A AUX HI and LO. 5.004 DISP Clamp the current clamp around the 50-turn coil. 5.005 HEAD AC CURRENT: {300A Range} 5.006 5520 300 50.0A 1.5% 0.5U 60H SI T5 6.001 DISP Connect a Fluke 5520A/COIL to the 5520A 20A and AUX LO. 6.001 DISP Clamp the current clamp around the 50-turn coil. 6.002 5520 300 300.0A 1.5% 0.3U 60H SI HC T5 7.001 END